Skip to main content

Table 3 Effective dose

From: Comparison of 640-slice Aquilon ONE CT scanner’s measured dosimetric parameters with ICRP dose reference levels for head, chest and abdominal CT examinations

Body region

Patient category

Weighting factor W T

Effective dose E D (mGy)

ICRP DRL

Measured

Rel. dev., %

Head

Pediatric

0.01

6.15

4.07

−33.8

Adult

0.01

10.50

10.08

−4.0

Chest

Pediatric

0.12

60.00

8.06

−86.6

Adult

0.12

78.00

33.86

−56.6

Abdomen

Pediatric

0.12

45.00

11.16

−75.2

Adult

0.12

93.60

42.42

−54.4

  1. Rel. dev. relative deviation from reference DRLs, s.d. standard deviation