Table 2 ICRP DRLs and measured dosimetric parameters for pediatric and adult CT head, chest and abdominal scans

From: Comparison of 640-slice Aquilon ONE CT scanner’s measured dosimetric parameters with ICRP dose reference levels for head, chest and abdominal CT examinations

Body Region

Patient category

CTDIvol ± s.d (mGy)

DLP ± s.d. (mGy/cm)

p-value

ICRP DRL

Measured

Rel. dev. %

ICRP DRL

Measured

Rel. dev. %

 

Head

Pediatric

56.11

23.87 ± 10.18

−57.5

614.58

406.59 ± 10.18

−33.8

0.001

Adult

60.00

63.29 ± 10.18

5.5

1050.00

1008.26 ± 10.18

−4.0

0.001

Chest

Pediatric

30.00

1.70 ± 10.18

−94.3

500.00

67.20 ± 10.18

−86.6

0.002

Adult

30.00

5.92 ± 10.18

−80.3

650.00

282.14 ± 10.18

−56.6

0.001

Abdomen

Pediatric

30.00

2.70 ± 10.18

−90.0

375.00

93.00 ± 10.18

−75.2

0.001

Adult

35.00

6.79 ± 10.18

−80.6

780.00

353.48 ± 10.18

−54.7

0.001